Abstract

Several different materials, including one which was porous, were studied to assess their properties as pacemaker electrode tips. Leads were implanted in sheep for periods up to one year. Electrical measurements were made during the implant period and histopathological examination performed after sacrifice. Although titanium vapor-deposited carbon, and silver did not lower the chronic stimulation threshold below that of platinum, their electrical characteristics were within generally acceptable limits. Zinc evoked a severe tissue reactions and a high threshold. Porous titanium alloy electrodes demonstrated reduced dislodgement, more frequent attachment and a lower sensing impedance than other electrodes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.