Abstract

An investigation of the tissue reaction to five porous titanium alloy pacemaker electrode tips inserted in sheep hearts was made. Attachment of the electrode tips to the endocardial wall was facilitated by fibrous tissue ingrowth into the porous material. Vascularized fibrocollagenous tissue was found within the pores of the tips. One tip was found to be infected. The presence of occasional multinucleated foreign-body giant cells within the porosity was related to the particulate features of the porous tip. The electrode tips had a reduced sensing impedance compared with conventional solid platinum tips. The present findings are consistent with previous observations relating to porous cobalt-chromium alloy and porous platinum-iridium alloy tips and suggest that porous titanium alloy may be a suitable pacemaker electrode tip material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.