Abstract

The purpose of this paper is to evaluate the image quality and dose to patients in the radiography of sacroiliac joints and to perform a clinical comparative study of digital and conventional screen-film radiography. Routine radiography of sacroiliac joint was performed in 60 patients using digital and screen-film radiography. The visibility of five anatomical regions and the overall image quality were rated by experienced radiologists. Patient dose assessment in terms of entrance surface air kerma (ESAK) was performed. The digital system showed slightly improved visualisation of specific anatomical structures. Overall image quality was significantly better in the digital when compared with the screen-film imaging system. The average ESAK was 2.4 mGy in screen-film and 3.6 mGy in digital radiography. The digital radiography provided equal or better visibility of anatomical details and overall image quality, but on higher dose levels. Therefore, the practice on digital systems must be optimised.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.