Abstract

Potentiometric chloride sensors were fabricated by electrochemically chloridizing the surface of screen printed silver electrodes using amperostatic coulometry or asymmetric square wave voltammetry. In the case of coulometric chloridization, the effect of different chloride ion sources was also investigated. The response to chloride ions by each type of sensor was determined by measuring their potentials with respect to a commercial Ag/AgCl reference electrode over the chloride concentration range 1.8 to 3,545mg/L (approximately 50μM to 100mM). Results reveal that irrespective of fabrication method, all sensors exhibit near Nernstian responses with no significant relationships between fabrication and response parameters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.