Abstract

Small field dosimetry measurements including output factors are difficult due to lack of charged-particle equilibrium, occlusion of the radiation source, the finite size of detectors, and non-water equivalence of detector components. With available detectors significant variations could be measured that will lead to incorrect delivered dose to patients. IAEA/AAPM have provided a framework and formulation to correct the detector response in small photon fields. Monte Carlo derived correction factors for some commonly used small field detectors are now available, however validation has not been performed prior to this study. An Exradin A16 chamber, EDGE detector and SFD detector were used to perform the output factor measurement for a series of conical fields (5–30mm) on a Varian iX linear accelerator. Discrepancies up to 20%, 10% and 6% were observed for 5, 7.5 and 10 mm cones between the initial output factors measured by the EDGE detector and the A16 ion chamber, while the discrepancies for the conical fields larger than 10 mm were less than 4%. After the application of the correction, the output factors agree well with each other to within 1%. Caution is needed when determining the output factors for small photon fields, especially for fields 10 mm in diameter or smaller. More than one type of detector should be used, each with proper corrections applied to the measurement results. It is concluded that with the application of correction factors to appropriately chosen detectors, output can be measured accurately for small fields.