Abstract

An optically trapped prolate glass stylus is the force-sensing element of a novel scanning-force microscope. Stylus displacement is detected with the use of the forward scatter of the trapping laser beam. Radiation forces owing to the three-dimensional intensity distribution near the focus permit the stylus to be both positioned with fine control and oriented along the z (optic) axis. Details of 20-nm size appear in traces recorded with a crude stylus in a trap formed with 1064-nm radiation. The spring constant of the optical-force transducer is below 10−4 N/m, which is to be compared with ∼0.1 N/m for typical mechanical cantilevers used in atomic-force microscopy. This gentler technique should improve the sensitivity of scanning-force microscopy for the imaging of soft samples in aqueous media.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.