Abstract

Performance of a prompt gamma ray neutron activation analysis (PGNAA) setup depends upon thermal neutron yield at the PGNAA sample location. For a moderator, which encloses a sample, thermal neutron intensity depends upon the effective moderator volume excluding the void volume due to sample volume. A rectangular moderator assembly has been designed for the King Fahd University of Petroleum and Minerals (KFUPM) PGNAA setup. The thermal and fast neutron yield has been measured inside the sample cavity as a function of its front moderator thickness using alpha particle tracks density and recoil proton track density inside the CR-39 nuclear track detectors (NTDs). The thermal/fast neutron yield ratio, obtained from the alpha particle tracks density to proton tracks density ratio in the NTDs, shows an inverse correlation with sample to moderator volume ratio. Comparison of the present results with the previously published results of smaller moderators of the KFUPM PGNAA setup confirms the observation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.