Abstract

AbstractPerylene is strongly fluorescent as long as it is in solution. This has enabled visual observation of non‐evaporated sample material in a “transparent injector”, i.e. in a heated glass device imitating a conventional vaporizing injector.Three scenarios of sample evaporation are described. Some samples (solvents) are nebulized and “flash evaporated” in the gas phase between the needle exit and the column entrance (Scenario 1). With most solvents, the liquid leaves the syringe needle as a thin jet which rushes through both the empty vaporizing chamber and the split outlet at high velocity, often without substantial evaporation. It does not touch the surfaces of the insert and passes round the bend at the bottom of the device without any problem (Scenario 2). Some samples are splashed on to the insert wall, wet it, and evaporate rather slowly from this surface (Scenario 3).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.