Abstract

The strength–duration curve for tissue excitation can be modeled by a parallel resistor–capacitor circuit that has a time constant. We tested several short-duration electric generators: five electric fence energizers, the Taser X26 and a high-frequency generator to determine their current-versus-time waveforms. We estimated their safety characteristics using existing IEC and UL standards for electric fence energizers. The current standards are difficult to follow, with cumbersome calculations, and do not explicitly explain the physiological relevance of the calculated parameters. Hence we propose a new standard. The proposed new standard would consist of a physical RC circuit with a certain time constant. The investigator would discharge the device into a passive resistor–capacitor circuit and measure the resulting maximum voltage. If the maximum voltage does not exceed a limit, the device passes the test.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.