Abstract
Different defibrillation waveforms exhibit different efficacies. It has been proposed that a resistor-capacitor (RC) model of the myocardial membrane can explain these differences. Both the maximum capacitor voltage (charge banking) and the remaining voltage after the pulse (charge burping) have been proposed as efficacy predictors. In this paper, we test these hypotheses against 140 defibrillation waveforms. Data is reported from 159 guinea pigs. For each tested waveform, we calculated to within a fixed constant the minimum and maximum voltage on the capacitor, and the voltage remaining after the shock for time constants (/spl tau/) between 0.1 ms and 4.0 ms. Each parameter was tested to determine if it would be predicted to have a 10% impact on defibrillation efficacy. No parameter was a significant predictor of defibrillation efficacy for /spl tau/>2.0 ms. The maximum voltage was significant for short time constants (/spl tau/<=2.0 ms, p<0.005). For the very short time constants (<=0.5 ms) the voltage remaining was significant (p<0.05), and, for the shortest time constants (t=0.1 ms), the minimum voltage was a significant predictor (p<0.05).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.