Abstract
The usage and safety of the Boost and Ease-off features in the CamAPS FX hybrid closed-loop system were analyzed in a retrospective analysis of real-world data from 7,464 users over a 12-month period. Boost was used more frequently than Ease-off, but for a shorter duration per use. Mean starting glucose was above range for Boost (229 ± 51 mg/dL), and within range for Ease-off (114 ± 29 mg/dL). Time spent below 70 mg/dL was low during Boost periods [median (interquartile range; IQR) 0.0% (0.0, 0.5%)], and lower than during no Boost periods [2.1% (1.2, 3.4%)], while time spent above 180 mg/dL was lower during Ease-off periods (15 ± 14%) compared with no Ease-off periods (25 ± 12%). There were no episodes of severe hypoglycemia or diabetic ketoacidosis attributed to Boost or Ease-off use. Boost and Ease-off allow users to engage safely with CamAPS FX to manage their glucose levels during periods of more-than-usual and less-than-usual insulin needs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.