The safety and efficacy of high-pressure injection through a stopcock was tested using a disposable flow switch model. Multiple catheters, contrast agents, injection rates, and pressure combinations were used. These demonstrated that, after insuring the adequacy of the stopcock connections, there will be no leakage, decrease in injection rate, or increase in required pressure when injecting through the flow switch stopcock even at levels appropriate for aortic and pulmonary arteriography.
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.