Abstract

BackgroundThe role of totally implantable central venous port (TICVP) system is increasing. Implantation performed by radiologist with ultrasound-guided access of vein and fluoroscope-guided positioning of catheter is widely accepted nowadays. In this article, we summarized our experience of TICVP system by surgeon and present the success and complication rate of this surgical method.Materials and MethodsBetween March 2009 and December 2010, 245 ports were implanted in 242 patients by surgeon. These procedures were performed with one small skin incision and subcutaneous puncture of subclavian vein. Patient's profiles, indications of port system, early and delayed complications, and implanted period were evaluated.ResultsThere were 82 men and 160 women with mean age of 55.74. Port system was implanted on right chest in 203, and left chest in 42 patients. There was no intraoperative complication. Early complications occurred in 11 patients (4.49%) including malposition of catheter tip in 6, malfunction of catheter in 3, and port site infection in 2. Late complication occurred in 12 patients (4.90%).ConclusionSurgical insertion of TICVP system with percutaneous subclavian venous access is safe procedures with lower complications. Careful insertion of system and skilled management would decrease complication incidence.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.