Abstract
Standards for safe use of lasers like those of the American National Standard Institute (ANSI) do not give specific guidelines regarding the safety norms for a Maxwellian view laser scanning device (for example, the scanning laser ophthalmoscope). In this note it is shown that a pulsed extended source ocular exposure system, which is described by the ANSI standards, can be used as a conservative model of such a laser scanning device. With some adjustments to this model, the Maxwellian character can be included and the maximum permissible corneal power also can be normalized independently of the field size.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.