Abstract

Summary. Linear gradient HPLC on a C8 column has been used for separation of individual related substances of amoxicillin listed in the European Pharmacopoeia and a newly identified degradation impurity. The USP plate count for the amoxicillin peak was more than 3000 and USP tailing for the same peak was less than 2.0. Forced degradation studies were conducted on amoxicillin drug substance using ICH stress study guidelines to demonstrate the specificity and stability-indicating nature of the method. A new impurity observed after thermal and alkaline degradation was identified as N-pivaloylamoxicillin. The LOD and LOQ for individual related substances were below 0.045 and 0.086% (w/w), respectively. The method was fully validated in accordance with ICH analytical method validation guidelines. The results of the study prove the method is specific, precise, linear, robust, and can be used for evaluation of the stability of amoxicillin drug substance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.