Abstract

To gather initial procedural experiences with the ThromCat XT, a new rotational thrombectomy catheter primarily developed for coronary interventions. The ThromCat XT is a 150-cm rapid exchange thrombectomy device consisting of an atraumatic tip and a steel helix with a kink-resistant covering. It was employed in native arteries, veins, grafts and stents in ten patients presenting with acute and subacute thrombembolic occlusions. Technical success with a restoration of flow was achieved in 70% of patients. The mean thrombectomy time was 8.0±2.33 min, and the aspirated blood volume ranged from 120 mL to 280 mL. Peripheral thrombembolism was detected in two cases, and embolic protection was applied in four cases. Vessel injuries and catheter failures were not observed in any of the cases. The ThromCat XT is an easy-to-handle, reliable and atraumatic device for the removal of fresh thrombi in native and artificial vessels. In our series, the thrombus age-especially if greater than five days-had a major impact on technical success.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.