Abstract
Accurate calculation of set-up margin is a prerequisite to arrive at the most optimal clinical to planning target volume margin. The aim of this study was to evaluate the compatibility of different on-board and in-room stereoscopic imaging modalities by calculating the set-up margins (SM) in stereotactic body radiotherapy technique accounting and unaccounting for rotational positional errors (PE). Further, we calculated separate SMs one based on residual positional errors and another based on residual + intrafraction positional errors from the imaging data obtained in a dual imaging environment. A total of 22 lung cancer patients were included in this study. For primary image guidance, four-dimensional cone beam computed tomography (4-D CBCT) was used and stereoscopic ExacTrac was used as the auxiliary imaging. Following table position correction (TPC) based on the initial 4-D CBCT, another 4-D CBCT (post-TPC) and a pair of stereoscopic ExacTrac images were obtained. Further, during the treatment delivery, a series of ExacTrac images were acquired to identify the intrafraction PE. If a, b and c were the observed translational shifts in lateral (x-axis), longitudinal (y-axis) and vertical direction (z-axis) and α, β and γ were the rotational shifts in radians about the same axes, respectively, then the resultant translational vectors (A, B and C) were calculated on the basis of translational and rotational values. Set-up margins were calculated using residual errors post-TPC only and also using intrafraction positional errors in addition to the residual errors. Residual and residual + intrafraction SM were calculated from a dataset of 82 CBCTs and 189 ExacTrac imaging sessions. CBCT-based mean ± SD shifts in translational and rotational directions were 0.3 ± 1.8mm, 0.1 ± 1.8mm, - 0.4 ± 1.6mm, 0.1 ± 0.4°, 0.0 ± 1.0° and 0.3 ± 0.7°, respectively, and for ExacTrac - 0.1 ± 1.8mm, 0.2 ± 2.4mm, - 0.6 ± 1.8mm, 0.1 ± 1.2°, - 0.2 ± 1.3° and - 0.1 ± 0.6°, respectively. Residual SM without considering the rotational correction in x, y and z directions were 5.0mm, 4.5mm and 4.4mm; rotation-corrected SM were 4.4mm, 4.0mm and 5.5mm, respectively. Residual plus intrafraction SM were 5.5mm, 6.6mm and 6.2mm without considering the rotational corrections, whereas they were 5.0mm, 6.3mm and 6.2mm with rotational errors accounted for. Accurate calculation of set-up margin is required to find the clinical to planning target volume margin. Primary and auxiliary imaging margins fall in the range of 4.0 to 5.5mm and 5.0 to 7.0mm, respectively, indicating a higher SM for X-ray-based planar imaging techniques over three-dimensional cone beam images. This study established the degree of mutual compatibility between two different kinds of widely used set-up imaging modalities, on-board CBCT and in-room stereoscopic imaging ExacTrac. It also describes the technique to calculate the residual and residual plus intrafraction SM and its variation in a dual imaging environment accounting for rotational PE in stereotactic body radiotherapy of lung.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.