Abstract

An adaptive Kalman filter-based fusion algorithm capable of estimating respiration rate for unobtrusive respiratory monitoring is proposed. Using both signal characteristics and a priori information, the Kalman filter is adaptively optimized to improve accuracy. Furthermore, the system is able to combine the respiration-related signals extracted from a textile ECG sensor and an accelerometer to create a single robust measurement. We measured derived respiratory rates and, when compared to a reference, found root-mean-square error of 2.11 breaths-per-minute (BrPM) while lying down, 2.30 BrPM while sitting, 5.97 BrPM while walking, and 5.98 BrPM while running. These results demonstrate that the proposed system is applicable to unobtrusive monitoring for various applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.