Abstract

The separation of telechelic poly(methyl methacrylate) (PMMA) prepolymers based on the number of end-groups under critical liquid chromatography (LC) conditions has been studied using a bare-silica column, which can interact with polar functional groups. The critical solvent compositions for non-functional, mono-functional and bi-functional PMMAs were determined in normal-phase LC using mixtures of acetonitrile and dichloromethane (DCM) of varying composition as the mobile phase. The telechelic prepolymers were successfully separated according to hydroxyl (OH) functionality (with zero, one, or two OH groups, respectively) under the critical conditions, in which fast (5 min), base-line separations were obtained independent of molecular weight. Changing the column temperature, flow rate, and mobile-phase composition within a certain range did not affect the functionality separation. Therefore this isocratic LC separation method is quite robust. Evaporative light-scattering detector (ELSD) calibration curves were used for the quantitative analysis of functional PMMA prepolymers.