Abstract

The concept of tissue-maximum ratios (TMR) as a basis of dose computations has posed problems when applied to a whole range of clinically used megavoltage beams. Another problem is the definition of scatter-maximum ratio (SMR) which assumes a value of zero at the depth of maximum dose. This paper describes new methods of measuring collimator and phantom scatter correction factors. The definitions of TMR and SMR are modified in order to compute phantom scatter at any depth, including depth of maximum dose. The revised concept is basic and general enough that it can be applied to x-ray beams of any energy fields of any shape and isocentric as well as non-isocentric modes of treatment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.