In this study, we developed a simple and sensitive reversed phase ion-pair chromatographic method for the analysis of C4-C6 sugar alcohols. The method is based on the on-line complexation of sugar alcohols with molybdate ion. The resulting dinuclear anionic complexes can be separated on a reversed-phase C18 column using tetrabutylammonium chloride as an ion-pairing reagent. The mobile phase (pH 3.1) consisted of 0.1 mM disodium molybdate, 1 mM hydrochloric acid and 0.4 mM tetrabutylammonium chloride – 10% v/v methanol. By complexing with molybdate ion, sugar alcohols can be detected by their UV absorption at 247 nm with high resolution and sensitivity. The quantification limits of the examined sugar alcohols calculated at S/N = 10 were 0.1 mM for erythritol and xylitol and 0.01 mM for arabitol, sorbitol, mannitol and dulcitol. The detector response was linear over three orders of magnitude of sugar alcohol concentration. The proposed method was successfully applied to measure sugar alcohols in health drinks, eyedrops and mouthwashes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.