Abstract

To evaluate the use of endobronchial forceps to retrieve tip-embedded inferior vena cava (IVC) filters. This institutional review board-approved, HIPAA-compliant retrospective study included 114 patients who presented with tip-embedded IVC filters for removal from January 2005 to April 2014. The included patients consisted of 77 women and 37 men with a mean age of 43 years (range, 18-79 years). Filters were identified as tip embedded by using rotational venography. Rigid bronchoscopy forceps were used to dissect the tip or hook of the filter from the wall of the IVC. The filter was then removed through the sheath by using the endobronchial forceps. Statistical analysis entailed calculating percentages, ranges, and means. The endobronchial forceps technique was used to successfully retrieve 109 of 114 (96%) tip-embedded IVC filters on an intention-to-treat basis. Five failures occurred in four patients in whom the technique was attempted but failed and one patient in whom retrieval was not attempted. Filters were in place for a mean of 465 days (range, 31-2976 days). The filters in this study included 10 Recovery, 33 G2, eight G2X, 11 Eclipse, one OptEase, six Option, 13 Günther Tulip, one ALN, and 31 Celect filters. Three minor complications and one major complication occurred, with no permanent sequelae. The endobronchial forceps technique can be safely used to remove tip-embedded IVC filters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.