Abstract

Removing a retrievable inferior vena cava (IVC) filter can be extremely difficult with the use of standard techniques if the filter is tilted and embedded in the wall of the IVC. The use of rigid endobronchial forceps has been described in case reports to remove embedded IVC filters, and the present report describes the use of this technique to remove a series of tip-embedded IVC filters in two separate institutions. The medical records were reviewed of 21 consecutive patients at two institutions who underwent attempted IVC filter removal with rigid endobronchial forceps over a 34-month period. The mean age of patients was 32.4 years (range, 14.1-54.1 y). The patients had the following filters: Recovery (n = 6), G2 (n = 10), Günther Tulip (n = 4), and OptEase (n = 1). Rotational or biplane venography was used to confirm that the filters were tilted and embedded in the wall of the IVC in all 21 patients. Rigid endobronchial forceps were used successfully to remove 20 embedded IVC filters in 21 patients. There was one case of failure to remove an embedded suprarenal G2 filter. There were no major complications. Rigid endobronchial forceps may be used as a reliable option for removal of embedded IVC filters.