Abstract

AbstractBasic expressions are derived for both the retention time and the effective separation factor in serially coupled GC columns. The retention time is determined by two main parameters. The first is the fractional time spent by an unretarded solute in each column which, in turn, is determined by the relative column lengths and flow velocities through each column. The second parameter is the relative mass distribution coefficient of a particular solute in each column; a variable that can be adjusted by changing the relative temperatures of the columns. The expression for the effective separation factor relates the measured separation factor for the series combination to the separation factors on the individual columns, the fractional time spent by an unretarded peak in each column, as well as the relative values of the mass distribution coefficients of a particular solute on the different columns.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.