Abstract
This paper presents the results of an intercomparison for extremity dosemeters organised by the European radiation dosimetry group in 2009. In total, 59 systems were tested during this exercise including ring, stall and wrist dosemeters. A total of 1652 dosemeters were irradiated in the selected fields of photons and beta radiation qualities on appropriate phantoms (ISO finger and pillar phantom) in the dose quantity H(p)(0.07). All irradiations were carried out in selected accredited reference dosemetry laboratories (Seibersdorf Laboratories, Austria and IRSN, France). The results show that, especially at low-energy beta radiations ((85)Kr) and for beta irradiations with large angles of incidence (60°), many tested systems show pronounced under responses. On the other hand, for photon irradiations down to energies of 16 keV most systems showed good results. A participants meeting was held at IM2010 with discussion on both general aspects of this intercomparison and specific problems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.