Abstract
An intercomparison of spectral irradiance measurements by 12 national laboratories has been carried out between 1987 and 1990. The intercomparison was conducted under the auspices of the Comité Consultatif de Photometrie et Radiometrie (CCPR) of the Comité International des Poids et Mesures, and the National Institute of Standards and Technology (NIST) served as the pilot laboratory. The spectral range of the intercomparison was 250 to 2400 nm and the transfer standards used were commercial tungsten-halogen lamps of two types. The world-wide consistency of the results (one standard deviation) was on the order of 1% in the visible spectral region and 2 to 4% in the ultraviolet and infrared portions of the spectrum. The intercomparison revealed no statistically significant differences between spectral-irradiance scales based on blackbody physics and absolute detector radiometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Research of the National Institute of Standards and Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.