Abstract

GAGG(Ce) is a novel scintillator with a fast response and high light output without a hygroscopic nature. It is expected to be a useful detector for charged particles at high-counting rates. However, the response of the GAGG(Ce) scintillator to charged particles has not been fully examined. In the present work, the light output and energy resolution of the GAGG(Ce) scintillator were measured for protons and alpha particles at E p=5–68 MeV and E α=8–54 MeV as well as gamma rays at E γ=662 keV from a 137Cs source. The results were compared with those of the CsI(Tl) scintillator. The scintillation efficiencies dL/dE of the GAGG(Ce) and CsI(Tl) scintillators were obtained and parametrized as a function of linear energy transfer dE/dx.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.