Abstract

We report measurements of the response of large-volume superheated droplet detectors for the SIMPLE project to filtered monochromatic neutron beams of energies 54 and 149 keV. The results continue to indicate a significantly higher than anticipated response at low temperatures, but provide good support for the theoretical device calibration in temperature. A significant moderation of the incident beams by the aqueous gel of the detector and the temperature-stabilizing water bath is observed, such that the detector response is primarily governed by the superheating factor of the detector refrigerant. Translated into background neutron sensitivity of WIMP searches, these features indicate that a good choice of operating conditions is 9°C and 2 atm, with the temperature implying a mass sensitivity >5 GeV given that sufficient background reduction below this temperature can be achieved.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.