Abstract
Respiratory rate is among the first vital signs to change in deteriorating patients. The aim was to investigate the agreement between respiratory rate measurements by three different methods. This prospective observational study included acutely admitted adult patients in a medical ward. Respiratory rate was measured by three methods: a standardised approach over 60 s while patients lay still and refrained from talking, by ward staff and by a wireless electronic patch (SensiumVitals). The Bland-Altman method was used to compare measurements and three breaths per minute (BPM) was considered a clinically relevant difference. We included 50 patients. The mean difference between the standardised approach and the electronic measurement was 0.3 (95% CI: -1.4 to 2.0) BPM; 95% limits of agreement were -11.5 (95% CI: -14.5 to -8.6) and 12.1 (95% CI: 9.2 to 15.1) BPM. Removal of three outliers with huge differences lead to a mean difference of -0.1 (95% CI: -0.7 to 0.5) BPM and 95% limits of agreement of -4.2 (95% CI: -5.3 to -3.2) BPM and 4.0 (95% CI: 2.9 to 5.0) BPM. The mean difference between staff and electronic measurements was 1.7 (95% CI: -0.5 to 3.9) BPM; 95% limits of agreement were -13.3 (95% CI: -17.2 to -9.5) BPM and 16.8 (95% CI: 13.0 to 20.6) BPM. A concerning lack of agreement was found between a wireless monitoring system and a standardised clinical approach. Ward staff's measurements also seemed to be inaccurate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.