Abstract

Introduction Spirometry allows an indirect control of moving targets in breathing hold (BH) or their monitoring in freebreathing (FB). FB technique requires a synchronization between the CT-scan acquisition (4D-CT) and the breathing cycles. This technique is not included in every system. The Slow-CT scan is a possible alternative to optimize the monitoring of moving targets. The purpose of this study is to evaluate the feasibility of non-synchronized FB using Slow-CT scan in terms of image quality, breathing amplitude and respiratory rate. Methods Standard-CT, Slow-CT and 4D-CT acquisitions are made with a Siemens Somatom Scope Power scan, with a voltage of 130 kV and modulated intensity (CARE Dose 4D®). UH, low contrasts, slice thickness and spatial resolution are analyzed on Catphan® 504. The moving target, a lung equivalent cylindrical insert including a 3 cm diameter sphere, is set in motion by Quasar™ Respiratory Motion. The breathing cycles are created by Trace Generator. Reconstruction performances are evaluated for breaths from 10 to 30 breaths per minute (BPM) and an amplitude of 2 cm. The radiation dose is estimated by the volumetric computed tomography dose index (CTDI). The choice of the FB amplitude margins is made from a retrospective analysis of 15 patients treated in BH conditions. The image quality of the 4D-CT Average and Slow-CT sequences is examined on patients. Results Noise is reduced up to 30% in Slow-CT (σ = 3.0 UH) in comparison with 4D-CT Average (σ = 4.9 UH). A systematic error of +0.35 mm of slice thickness reconstruction occurs for 4D-CT Average. The low contrasts score (visual) increases in 4D-CT Average. In 4D-CT Average (>50 s), the amplitude reconstruction is faithful (49.8 mm instead of 50.0 mm in theory) with the 10 BPM/2 cm amplitude couple. In Slow-CT (>60 s), for BPM between 12 and 18, the mean reconstructed amplitude is 47 mm. It is 41 mm for BPM ⩽10 or ⩾30. The amplitude reconstruction is optimized for 15 BPM (49.5 mm). In Standard-CT (6–20 s), it appears distortions in the 3 reconstruction planes and relatives errors up to 60 %. In Slow-CT and 4D-CT, the mean CTDI are respectively 16.4 mGy and 15.2 mGy. The optimized Slow-CT acquisition parameters are 130 kV, collimation of 16x0.6 mm, 2 mm reconstructed slice, pitch of 0.4, rotation time of 1.5 s and a ratio (Acquisition time (s))/BPM⩾4. The retrospective analysis of patients shows a maximum interfraction amplitude difference of 100 mL. The spirometer calibration includes a 60 mL uncertainty (2% of 3 L). A margin of 100 mL is chosen as upper limit (inhalation); the baseline, allowing an automatic stabilization of the system display, is the lower limit (exhalation). On clinical images, Slow-CT shows a motion blur reduction (heart, mediastinum) and a noise reduction in a range of 10–15%. Conclusions The Slow-CT sequence offers an alternative to 4D-CT for incompatible systems. Parameters such as collimation, pitch, rotation time and the ratio (Acquisition time (s))/BPM⩾4 must be adapted. CTDI are similar for Slow-CT and 4D-CT. Margins optimization and further clinical cases are ongoing. A study about dose optimization while obtaining the same amount of signal will be carried out using different level of noise reduction algorithm.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call