Abstract
In order to solve the unstable problem of conventional methods for measuring resonant frequency of quartz crystal, a method based on DDS (Direct Digital Synthesis) and real-time peak searching is proposed to increase stability of measurement. By actively generating excitation signal using DDS, the response of quartz crystal can be obtained even under heavy load. Frequency response of impedance is studied to reveal the relationship between minimum impedance and resonant frequency. Electric circuit and real-time searching algorithm is designed to track resonant frequency of quartz crystal. Experimental results show when parameter of quartz crystal changes 392.0%, the resonant frequency could still be measured by the developed method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.