Abstract

Nuclear Medicine is a modality that has been growing a lot in Brazil and in the world, bringing new radiopharmaceuticals and technologies for the diagnosis and treatment of diseases, making them more effective and accurate. In the development of new radiopharmaceuticals for use in positron emission tomography (PET), the preclinical studies step largely uses PET scanners dedicated to small animals. To achieve this, quality control tests must be carried out to ensure the efficiency of the equipment. Thus, this work aimed to evaluate the spatial resolution of the small animals PET scanner of the CDTN using the isotope 68Ga. In this sense, a microPET hot rod phantom and a commercial 68Ge/68Ga generator were used to obtain PET images. The analysis of the 68Ga-PET images of the simulator was carried out qualitatively and quantitatively and revealed that the spatial resolution of the system using the 68Ga is 3.5 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.