Abstract

A new hyper-fast gas chromatography method with less than 90 s runtime including the column cool down was developed for the analysis of four gases and 16 residual solvents, combining a CO2 cryofocusing with a flow-field thermal gradient gas chromatograph (FF-TG-GC) and ToF-MS. The extremely low analysis time can be achieved by combining the new FF-TG-GC and a very short Rxi-624 Sil MS separation column with a small inner diameter and small film thickness (2.05 m × 0.1 mm × 1.0 µm).The column is inserted into a low thermal mass, resistively heated stainless steel capillary. This enables fast temperature programs with heating rates up to 3000 °C/min and a column cool down within a few seconds. In addition to temporal temperature gradients, the FF-TG-GC can generate a spatial temperature gradient that leads to an improved peak shape. Further, an external liquid CO2 cryo-trap was designed in order to reduce the injection bandwidths of analytes and to take full advantage of the resolving power of the separation column. No modifications are required to the FF-TG-GC for the use of the cryogenic trap, as the cooled spot is heated by the resistively heated stainless steel capillary during the temperature program. With cryofocusing, analyzed residual solvents are baseline separated. R2 values over 0.99 for calibration curves and low relative standard deviations (mainly < 3%) for repeatability tests were obtained.