Abstract

With the increasing demand for stable, accurate and reliable frequency standard sources in the field of measurement and control, rubidium atomic clocks with strong environmental adaptability and low price are widely used but are limited by long-term frequency drift. In this paper, 1PPS signal as the reference signal provided by the National Time Service Center, and a rubidium atomic clock frequency automatic calibration system was designed and implemented. Experimental results show that, the clock difference is less than 8ns between the rubidium atomic clock and the master clock of the NTSC, the Modified Allan Deviation is 1.83E-12@100s, and 2.70E-13@10000s. The experiment proves that the self-calibrating frequency system gives rubidium atomic clock a good accuracy and long-term stability.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call