Abstract

For the purpose of realizing the dynamic angle measurement and calibration of the inertial equipment (such as rotary table), a dynamic angle measurement system based on the autocollimator and the fiber optic gyroscope (FOG), which can be used as an angle standard device, is designed in this paper. In this system, the scale factor of the FOG can be calibrated in the field application based on the autocollimator with the precision of 1 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">''</sup> in the range of ±0.05°. Furthermore, the angle measurement deviation between the FOG and the rotary table can be limited in ±3 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">''</sup> when the scale factor of the FOG is calibrated by the autocollimator when necessary. This research can guarantee the long-term accuracy of the scale factor of the FOG in the field application and it will greatly promote the applications of the FOG as an angle standard device in the field of dynamic angle measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call