Abstract
Summarized the research status of digital holography microscopy technology. Aiming at the experimental parameters which influences the quality of hologram recurrent image, the recording and reconstruction process of hologram are analyzed, the reconstruction distance are calculated which based on the reflective off-axis digital holographic light path. The experiment records the hologram of standard resolution board and reconstruct it under different light intensity ratio. The results show that the object and reference light intensity ratio have a significant impact on the quality of hologram recurrent image. Besides, it’s the key to determining an appropriate recording distance and object and reference light intensity ratio of improving the quality of hologram. The result would provide theoretical basis and technical conditions for studying the reflective digital holography based on 3-D reconstruction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IOP Conference Series: Materials Science and Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.