Abstract

The National High Magnetic Field Laboratory (NHMFL) has identified as one of its next high-field persistent-magnet targets a 21 T superconducting magnet system to be used in a Fourier Transform Ion Cyclotron Resonance (FT-ICR) Mass Spectrometer. The required fundamental parameters of a 21 T central field and 110 mm warm bore diameter are similar to those already successfully demonstrated in the NHMFL's Ultra-Wide Bore 900 MHz Nuclear Magnetic Resonance magnet system. However, there are several ancillary requirements of the FT-ICR spectrometer that require technical attention such as fringe field restrictions requiring shielding, length of uniformity zone, magnitude of inhomogeneity, horizontal orientation of the magnet axis and a limitation on the length of the cryostat. This paper describes the overall system requirements, features and status of the superconducting magnet conceptual design proposed for meeting these requirements