Abstract
The reproducibility of repeated single-voxel 1H MRS (SV-MRS) and spectroscopic imaging (MRSI) measurements of intramyocellular lipid (IMCL) in the tibialis anterior muscle of five lean and five overweight female Caucasians, during 7 days of controlled dietary fat and calorie intake, was assessed at 1.5 T. Duplicate measures of IMCL relative to total muscle creatine (IMCL/tCr) obtained 3 days apart by both SV-MRS and MRSI correlated well (r = 0.65 and r = 0.95, respectively, P < 0.05). The coefficients of variation for repeated measures of IMCL/tCr by SV-MRS and MRSI were 24.4% and 10.7%, respectively. IMCL/tCr measured by MRSI was higher in overweight subjects than in lean subjects (8.3 +/- 3.8 vs 4.3 +/- 2.4, P < 0.05). Although both methods achieved good reproducibility in measuring IMCL in vivo, MRSI was found to offer greater flexibility and reliability, and higher sensitivity to IMCL differences, whereas SV-MRS was advantageous with respect to shorter scan time and ease of implementation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.