Abstract
To provide faster laboratory data reporting, we evaluated the accuracy of Sysmex XN (Sysmex Inc, Kobe, Japan) absolute neutrophil count (ANC) in the presence of analyzer flagging. Sysmex XN and manual microscopy ANC were compared with 80 autovalidated control specimens and with 280 study specimens with analyzer flagging regarding immature granulocytes (IG) >3% or other leukocyte abnormalities. Specimens with ambiguous neutrophil clusters were excluded. A slight positive overall method bias was seen for Sysmex XN compared to manual microscopy (n = 280), 0.025 (95% confidence interval [CI], -0.023 to 0.069) × 109/L. With IG > 10% (n = 123) the bias was larger, but not clinically significant, 0.17 (95% CI, 0.060-0.25) × 109/L. No clinically significant difference was seen in neutropenic (ANC < 1.5 × 109/L) specimens (n = 91), 0.070 (95% CI, -0.013 to 0.14) × 109/L. These data indicate that Sysmex XN ANC can be reported in the presence of certain analyzer flagging to improve patient care.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.