Repetitive processing of air samples for SO/sub 2/(g) determination in a continuous-flow system is presented. Determination is based on the reaction of SO/sub 2/ with di-..mu..-hydroxo-bis(bis(1,10-phen anthroline)iron(III)), the dimer of ferriin, at a gas-liquid interface. The performance of an open tube coll reactor, a single-bead string reactor, and a midget impinger setup is discussed and compared. The proposed methodology allows determination of SO/sub 2/ in the 0.5 to 15.0 ppM (v/v) range with relative standard deviations of 2 to 4% (for 10 repetitive determinations). The main reagent was electrochemically regenerated by controlled potential electrolysis allowing the use of a closed-loop system. The number of determinations per hour are 20 with the midget impinger, 40 with the open tube coll reactor, and 50 with the single-bead string reactor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.