Abstract
Two methods are proposed for the determination of regional concentrations of glutamate in the rat brain as well as in human serum. Glutamate oxidase was immobilized on non-porous glass beads and glutamate dehydrogenase was immobilized on glass derivatives. These supports were employed for the construction of Single Bead String Reactors and Packed Bed Reactors, respectively, which in turn were linked to Flow Injection Analysis systems with either photometric or fluorometric detection. Analytical working curves are linear in the range 1-200 mumol/l for packed bed reactors and 10-500 mmol/l for single bead string reactors. The samples were pretreated depending on their origin and the applied measuring system. Optimal dilution factors were established for the two techniques. Optimal dilution ratios were established and the influence of several added substances was investigated. Recovery and method comparison studies including high performance liquid chromatography verified the accuracy of the proposed methods. Results from within-day and between-day measurements gave relative standard deviations of 4.7 and 5.9% for serum samples and 2.5 and 4.0% for brain samples, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.