Abstract

To investigate the repeatability of three-dimensional (3-D) cephalometric measurements for the clinical application of 3-D cephalometry. Forty-nine measurements that widely used in traditional cephalometric analyses were defined in 3-D cone-beam CT (CBCT) images. Three examiners identified landmarks on CBCT images of 17 subjects with normal occlusion, respectively, and 3-D measurements were exported automatically by software SimPlant. Inter-examiner reliability correlation coefficients (ICC) were obtained for all measurements. Repeatability of 36 measurements was high (ICC value greater than 0.9), including SNA, SNB. Repeatability of 11 measurements was moderate (ICC value between 0.8 and 0.9), including CoL-GoL, CoL-MSP. Repeatability of 2 measurements was low (ICC value lower than 0.8), including Gn-MSP and MPR-MSP. Most 3-D cephalometric measurements based on CBCT had high repeatability. However, some 3-D cephalometric measurements had limited repeatability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.