Abstract

Meibomian gland contrast may be a potential indicator of gland health, especially among isotretinoin users. We aimed to develop a repeatable and reliable method for measuring Meibomian gland contrast from meibography images. Lower lid (LL) and upper lid (UL) meibography were captured with the OCULUS Keratograph 5M (OCULUS, Inc) at 2 visits under the following 4 conditions: face centered with room lights on (C), left-turned face (L), right-turned face (R), and face centered with room lights off (CLO). Contrast was measured with Fiji (v2.0.0-rc-59). Coefficient of repeatability and limits of agreement (LOA) were determined using Bland-Altman plots. A total of 512 meibography images from 16 subjects (age ± SD = 24.8 ± 5.2 years; 13 female patients) were collected. Coefficient of repeatability between visits was 10.5 for UL and 14.9 for LL. Lower and upper LOA, respectively, for UL, compared with condition C, were -10.9 [95% confidence interval (CI), -13.5 to -8.3] and 6.2 (95% CI, 3.6-8.8) for L; -11.0 (95% CI, -13.8 to -8.1) and 7.0 (95% CI, 4.2-9.8) for R; and -9.0 (95% CI, -11.6 to -6.5) and 7.2 (95% CI, 4.7-9.8) for CLO. Lower and upper LOA, respectively, for LL, compared with condition C, were -18.1 (95% CI, -22.6 to -13.5) and 11.0 (95% CI, 6.5-15.5) for L; -15.3 (95% CI, -19.2 to -11.3) and 9.9 (95% CI, 6.0-13.9) for R; and -12.0 (95% CI, -15.1 to -8.8) and 8.2 (95% CI, 5.0-11.3) for CLO. Meibomian gland contrast is a repeatable and reliable measure for changes in Meibomian gland contrast greater than 11 in the UL and 18 in the LL.