Abstract

Mobile technology is increasingly used to measure visual acuity. Standards for chart-based acuity tests specify photometric requirements for luminance, optotype contrast and luminance uniformity. Manufacturers provide some photometric data but little is known about tablet performance for visual acuity testing. This study photometrically characterised seven tablet computers (iPad, Apple inc.) and three ETDRS (Early Treatment Diabetic Retinopathy Study) visual acuity charts with room lights on and off, and compared findings with visual acuity measurement standards. Tablet screen luminance and contrast were measured using nine points across a black and white checkerboard test screen at five arbitrary brightness levels. ETDRS optotypes and adjacent white background luminance and contrast were measured. All seven tablets (room lights off) exceeded the most stringent requirement for mean luminance (≥ 120 cd/m2) providing the nominal brightness setting was above 50%. All exceeded contrast requirement (Weber ≥ 90%) regardless of brightness setting, and five were marginally below the required luminance uniformity threshold (Lmin/Lmax ≥ 80%). Re-assessing three tablets with room lights on made little difference to mean luminance or contrast, and improved luminance uniformity to exceed the threshold. The three EDTRS charts (room lights off) had adequate mean luminance (≥ 120 cd/m2) and Weber contrast (≥ 90%), but all three charts failed to meet the luminance uniformity standard (Lmin/Lmax ≥ 80%). Two charts were operating beyond manufacturer’s recommended lamp replacement schedule. With room lights on, chart mean luminance and Weber contrast increased, but two charts still had inadequate luminance uniformity. Tablet computers showed less inter-device variability, higher contrast, and better luminance uniformity than charts in both lights-on and lights-off environments, providing brightness setting was >50%. Overall, iPad tablets matched or marginally out-performed ETDRS charts in terms of photometric compliance with high contrast acuity standards.

Highlights

  • The widespread use of mobile technology is generating innovative ways to improve health

  • The authors noted that luminance uniformity varied within and between devices, but with little detriment to contrast uniformity [12]. This current study aimed to develop this work for high contrast acuity testing

  • Two standards describe the photometric qualities required of visual acuity test charts [20,21]

Read more

Summary

Introduction

The widespread use of mobile technology is generating innovative ways to improve health. It is estimated that 500 million mobile device users will download healthcare applications (‘apps’) by 2015 [1]. This brings an opportunity to enhance detection of compromised vision. It is unsurprising that mobile technology, in particular the iPad tablet (Apple Inc., CA, USA), has been evaluated as an alternative to traditional chart-based methods for measuring vision [5,6,7,8,9,10,11,12,13,14]. Updated guidance regarding which applications and target platforms are appropriate for clinical use is needed, and regulation, accreditation and ‘kitemarking’ of various healthcare technologies including apps is planned in the UK [17]

Methods
Results
Discussion
Conclusion