Abstract

During normal breathing, the kidneys move up and down due to the diaphragm motion and the renal artery subsequently experiences bending. Whether bending of the renal artery during respiration impacts stent bending fatigue was not previously known. This study presents the first evaluation of the effect of the kidney movement on renal stent bending fatigue performance. Measurements of fluoroscopic images demonstrated a minor (1.7deg) change in bending angle between inspiration and expiration for an 18mm long renal stent after removing rigid-body motion. Finite element analysis revealed excellent fatigue resistance of the studied renal stent under simulated respiratory motion.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.