Abstract In physiotherapy, joint kinematics is conventionally analyzed using the universal mechanical goniometer involving manual measurements and is only possible for static assessments of biomechanical parameters. The aim of the present research is to estimate and validate the knee joint kinematic parameters using the inertial sensor in a wearable device. A video motion analysis study is performed using ‘Kinovea’ software for measurement validation. No statistical difference is observed in hypothesis testing (t-test, p>0.7). The Pearson correlation coefficient (r2) and intraclass correlation coefficient values are observed to be 0.99 and >0.99, respectively, for a 95% confidence interval in the inter-rater reliability test. Further, data normality using the ‘Kolmogorov-Smirnov’ test, and Bland-Altman analysis is performed to validate quantitative agreements. The Standard error measurement confirms the reliability between the two systems. The developed device can be used as an alternative to the universal mechanical goniometer to measure joint kinematics in dynamic movements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.