Abstract

The 2.2.2 cryptand [4,7,13,16,21,24-hexaoxa-1,10-diazabicyclo-(8.8.8)-hexacosane] was trapped efficiently on Dowex AG50W-X8 (100–200 mesh) cation exchange resin. The concentration of the 2.2.2 cryptand in water with 0.1–2% methanol, or in 1 N HCl, was decreased by a factor of >4000 using 1 mL (1.7 mequiv.) of AG50W-X8 and a flow rate of approx. 2 mL/min. K +, Cs +, Ag + and Ba 2+ forms were about equal in their ability to remove the 2.2.2 cryptand. A disposable cartridge containing 1 mL of hydrogen form resin was inserted into our automated 18FDG system so that the hydrolysate (in 2 cm 3 of 1 N HCl) would pass through the cartridge before final purification. No cryptand was detected in the final product as determined by TLC with idoplantinate visualization. The detection limit was 2.5 μg/mL. Less than 3% of the total starting radioactivity was retained by the cation column. Quality assurance tests including apyrogenicity, sterility, radiochemical purity, carbohydrate composition and pH were not compromised by the incorporation of the cryptand removal cartridge.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.