Abstract

A method is described for the removal of AC interference from the electrocardiogram in monitors which do not have isolated inputs. The patient's leads are connected to an isolation amplifier which provides isolation between the input and the output and has a high common mode rejection ratio. The output from the isolation amplifier is taken to the electrocardiographic monitor. The isolation amplifier not only provides additional patient safety but also reduces the level of AC interference without interfering with any frequency components in the electrocardiogram.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.