Abstract
Open-configuration magnetic resonance (MR) systems are becoming desirable for volume measurements of off-center samples due to their non-claustrophobic system configuration, excellent soft-tissue contrast, high efficiency, and low-cost. However, geometric distortion is produced by the unwanted background magnetic field and hinders volume measurements. The present study describes the characteristics of geometric distortion in off-center samples such as the thigh muscle and adipose tissue measurements using an open-type MR system. The American Association of Physicists in Medicine uniformity and linearity phantom was used for the detection and the evaluation of the geometric distortion. The geometric distortion decreased near the isocenter and increased toward the off-center. A cylindrical phantom image was acquired at the isocenter and was used as the distortion-free, reference image. Two cylindrical phantoms were scanned off-center at a position analogous to that of the human thigh. The differences between the two cylindrical phantom volumes and the reference volume were 1.62 % ± 0.16 and 5.18 % ± 0.14. Off-center-related MR imaging requires careful consideration for image interpretation and volumetric assessment of tissue.