Abstract

This study aimed to determine the absolute and relative reliability of concentric and eccentric flexion, extension, horizontal abduction, and adduction movements of the shoulder using a functional electromechanical dynamometer (FEMD). Forty-three active male university students (23.51 ± 4.72 years) were examined for concentric and eccentric strength of shoulder flexion, extension, horizontal abduction, and horizontal adduction with an isokinetic test at 0.80 m·s-1. Relative reliability was determined by intraclass correlation coefficients (ICCs) with 95% confidence intervals. Absolute reliability was quantified by the standard error of measurement (SEM) and coefficient of variation (CV). Reliability was very high to extremely high for all movements on concentric and eccentric strength measurements (ICC: 0.76-0.94, SEM: 0.63-6.57%, CV: 9.40-19.63%). The results of this study provide compelling evidence for the absolute and relative reliability of concentric and eccentric flexion, extension, horizontal abduction, and horizontal adduction shoulder isokinetic strength tests in asymptomatic adults. The mean concentric force was the most reliable strength value for all tests.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.